Beam focusing is one of the unique characteristics of ultrasonic phased array compared with conventional ultrasound.On the basis of two-dimensional radiated sound field of phased array,the three-dimensional radiated sound field was simulated in the paper,and then the effect of different frequencies,different number of array elements and different element spacings on focal spot,the depth of focus and the effect on horizontal and vertical resolution were analyzed.The optimal results of transducer parameters have certain reference value for the design of phased array probe.
Phase aberration correction for medical ultrasound systems has attracted a great deal of attention. Since phased array techniques are now widely employed for industrial non-destructive testing (NDT) applications in various fields, the problem of phase aberrations in the process of NDT testing is considered. The technique of cross-covariance for phase aberration correction is presented. The performance of the technique for phase aberration correction is tested by means of echo signals obtained in practical non-destructive testing experiment. The results show that the technique has the better accuracy of phase correction.